Alarm 23. No Gas
Alarm message ....................No Gas
Alarm sub message .............Connect ventilator
to gas
Priority of alarm ....................23
Monitor mode .......................All
Can alarm be muted ............No
Sounder Priority ...................High
Alarm Description:
If no gas is detected by the ventilator this alarm is
generated.
Action: Connect gas supplies to ventilator. If
generated whilst connected to a patient,
remove patient to alternative form of
ventilation.
Alarm 24. No O2 Supply
Alarm message ....................No O
2
Supply
Alarm sub message .............Connect oxygen
supply to ventilator
Priority of alarm ....................24
Monitor mode .......................All
Can alarm be muted ............No
Sounder Priority ...................High
Alarm Description:
If no O
2
is detected by the ventilator, this alarm is
generated.
Action: Connect gas supplies to ventilator. If
generated whilst connected to a patient,
remove patient to alternative form of
ventilation.
Alarm 25. No Air Supply
Alarm message ....................No Air Supply
Alarm sub message..............Connect air supply
to ventilator
Priority of alarm ....................25
Monitor mode .......................All
Can alarm be muted ............No
Sounder Priority....................High
Alarm Description:
If no Air is detected by the ventilator, this alarm is
generated.
Action: Connect gas supplies to ventilator. If
generated whilst connected to a patient,
remove patient to alternative form of
ventilation.
Alarm 26. Battery Fault
Alarm message ....................Battery Fault
Alarm sub message..............Internal battery not
connected
Priority of alarm ....................26
Monitor mode .......................All
Can alarm be muted ............No
Sounder Priority....................High
Alarm Description:
The battery can not be detected by the
ventilator’s monitoring system.
Action: Remove ventilator from service. The
ventilator will continue to ventilate on mains
power if connected to a patient.
Alarm 27. Battery Low
Alarm message ....................Battery Low
Alarm sub message..............Remove patient
from ventilator
Priority of alarm ....................27
Monitor mode .......................All
Can alarm be muted ............No
Sounder Priority....................High
Alarm Description:
The battery low signal from the power supply sub-
system is monitored while the system is running
on battery. When the signal is asserted an alarm
indicating imminent power failure is generated.
Action: Reconnect mains power if available. If
not available remove patient to alternative
form of ventilation immediately.