Alarm 28. Blocked Fresh Gas
Alarm message.................... Blocked Fresh Gas
Alarm sub message............. Fresh gas supply to
patient may be
blocked
Priority of alarm ................... 28
Monitor mode....................... All
Can alarm be muted ........... No
Sounder Priority................... High
Alarm Description:
If the fresh gas supply is blocked, this alarm is
generated.
Action: Check patient circuit.
Alarm 29. Leaking Fresh Gas
Alarm message.................... Leaking Fresh Gas
Alarm sub message............. Fresh gas supply to
patient may be
leaking
Priority of alarm ................... 29
Monitor mode....................... All
Can alarm be muted ........... No
Sounder Priority................... High
Alarm Description:
If the fresh gas supply is detected to be leaking,
this alarm is generated.
Action: Check patient circuit.
Alarm 30. System Fail (Serial Communication
Error)
Alarm message.................... System fail
Alarm sub message............. Monitor error
(comms)
Priority of alarm ................... 30
Monitor mode....................... Flow
Can alarm be muted ........... Yes
Sounder Priority................... High
Alarm Description:
Errors detected within the flow monitoring device.
Action: Remove patient to alternative form of
ventilation, then remove ventilator from
service.
Alarm 31. System Fail (Isolate System Error)
Alarm message.................... System Fail
Alarm sub message ............. Monitor isolated
system fail
Priority of alarm.................... 31
Monitor mode....................... Flow
Can alarm be muted ........... Yes
Sounder Priority ................... High
Alarm Description:
Errors detected within the flow monitoring device.
Action: Remove patient to alternative form of
ventilation, then remove ventilator from
service.
Alarm 32. Flow Monitor (Flow ADC Unable to
Calibrate)
Alarm message.................... System Fail
Alarm sub message ............. Unable to calibrate
flow ADC
Priority of alarm.................... 32
Monitor mode....................... Flow
Can alarm be muted ........... Yes
Sounder Priority ................... High
Alarm Description:
Errors detected within the flow monitoring device.
Action: Remove patient to alternative form of
ventilation, then remove ventilator from
service.
Alarm 33. Flow Monitor (Flow Sensor Defect)
Alarm message.................... Faulty Flow Sensor
Alarm sub message ............. Flow sensor
defective
Priority of alarm.................... 33
Monitor mode....................... Flow
Can alarm be muted ........... Yes
Sounder Priority ................... Medium
Alarm Description:
If one of the flow sensor wires break this alarm is
generated.
Action: Replace and discard the defective
flow sensor.