Alarm 46. Apnoea (Pressure)
Alarm message ....................Apnoea
Alarm sub message .............Period between
patient effort
exceeds Apnoea
limit
Priority of alarm ....................46
Monitor mode .......................Pressure
Can alarm be muted ............Yes
Sounder Priority ...................High
Alarm Description:
If a breath is not detected within a user specified
apnoea period, an apnoea alarm will be
generated. The proximal pressure is used to
detect a breath.
Action: Check patient. Check patient circuit.
Adjust breath trigger sensitivity.
Alarm 47. Main Power Fail
Alarm message ....................Main Power Fail
Alarm sub message .............Running on internal
battery
Priority of alarm ....................47
Monitor mode .......................All
Can alarm be muted ............Yes
Can alarm be disabled .........Yes
Sounder Priority ...................High
Alarm Description:
If the mains supply fails this alarm is generated.
Action: Check mains connection. Prepare
alternative form of ventilation whilst running
on battery.
Alarm 48. Fresh Gas Solenoid Fail
Alarm message ....................Fresh Gas Solenoid
Fail
Alarm sub message..............REMOVE
VENTILATOR
FROM SERVICE
Priority of alarm ....................48
Monitor mode .......................All
Can alarm be muted ............Yes
Sounder Priority....................High
Alarm Description:
When the ventilator pressure transducer detects
no gas from the solenoid, this alarm is generated.
Action: Remove patient to alternative form of
ventilation immediately, then remove
ventilator from service.
Alarm 49. Oxygen Cell Disconnected.
(Disconnected)
Alarm message ....................Oxygen cell
disconnected.
Alarm sub message..............The Oxygen cell is
not connected
Priority of alarm ....................49
Monitor mode .......................All
Can alarm be muted ............Yes
Sounder Priority....................Medium
Alarm Description:
A disconnection of the oxygen cell will generate
this alarm.
Action: Remove patient to alternative form of
ventilation, then remove ventilator from
service.