Alarm 50. Calibrate Oxygen Cell
Alarm message.................... Calibrate Oxygen
Cell
Alarm sub message............. The oxygen cell
needs calibrating
Priority of alarm ................... 50
Monitor mode....................... All
Can alarm be muted ........... Yes
Sounder Priority................... Medium
Alarm Description:
If at any time the measured oxygen is >100% this
alarm will be generated.
Action: Recalibrate O
2
sensor. If alarm re-
appears remove patient to alternative form of
ventilation, then remove ventilator from
service.
Alarm 51. Oxygen Cell Exhausted
Alarm message.................... Oxygen Cell
Exhausted
Alarm sub message............. A new oxygen cell is
required
Priority of alarm ................... 51
Monitor mode....................... All
Can alarm be muted ........... Yes
Sounder Priority................... Medium
Alarm Description:
If the oxygen cell could not be calibrated during
calibration this alarm will be generated.
Action: Remove patient to alternative form of
ventilation, then remove ventilator from
service.
Alarm 52. Oxygen Calibration Fail
Alarm message.................... Oxygen calibration
fail
Alarm sub message ............. Check Oxygen
supply
Priority of alarm.................... 52
Monitor mode....................... All
Can alarm be muted ........... Yes
Sounder Priority ................... Medium
Alarm Description:
If during 100% oxygen calibration the measured
oxygen level is below a certain level this alarm is
generated.
Action: Remove patient to alternative form of
ventilation, then remove ventilator from
service.
Alarm 53. High Oxygen Level
Alarm message.................... High Oxygen Level
Alarm sub message ............. The Oxygen level is
higher than set
Priority of alarm.................... 53
Monitor mode....................... All
Can alarm be muted ........... Yes
Sounder Priority ................... High
Alarm Description:
If the delivered oxygen differs from the user
selected level by more than 5%, the above alarm
is generated.
Action: Recalibrate O
2
sensor. If alarm re-
appears, remove patient to alternative form of
ventilation, then remove ventilator from
service.