Page 134
Alarm 5. 101 System Fail (Memory Checksum
Error)
Alarm message ....................Monitor EEPROM
fail
Alarm sub message .............Monitor checksum
Fail
Alarm code ...........................101
Priority of alarm ....................5
Monitor mode .......................All
Can alarm be muted ............No
Sounder Priority ...................High
Alarm Description:
At power up flow data in EEPROM corrupt.
Action: Remove ventilator from service.
Alarm 6. 102 System Fail (Memory Checksum
Error)
Alarm message ....................Monitor EEPROM
fail
Alarm sub message .............Monitor checksum
Fail
Alarm code ...........................102
Priority of alarm ....................6
Monitor mode .......................All
Can alarm be muted ............No
Sounder Priority ...................High
Alarm Description:
At power up oxygen data in EEPROM corrupt.
Action: Remove ventilator from service.
Alarm 7. 103 System Fail (Memory Checksum
Error)
Alarm message ....................Monitor EEPROM
fail
Alarm sub message .............Monitor checksum
Fail
Alarm code ...........................103
Priority of alarm ....................7
Monitor mode .......................All
Can alarm be muted ............No
Sounder Priority ...................High
Alarm Description:
At power up pressure offset data in EEPROM
corrupt.
Action: Remove ventilator from service.
Alarm 8. 104 System Fail (Memory Checksum
Error)
Alarm message ....................Monitor EEPROM
fail
Alarm sub message..............Monitor checksum
Fail
Alarm code ...........................104
Priority of alarm ....................8
Monitor mode .......................All
Can alarm be muted ............No
Sounder Priority....................High
Alarm Description:
At power up pressure gain data in EEPROM
corrupt.
Action: Remove ventilator from service.
Alarm 9. 105 System Fail (Memory Checksum
Error)
Alarm message ....................Monitor EEPROM
fail
Alarm sub message..............Monitor checksum
Fail
Alarm code ...........................105
Priority of alarm ....................9
Monitor mode .......................All
Can alarm be muted ............No
Sounder Priority....................High
Alarm Description:
At power up pressure time constant data in
EEPROM corrupt.
Action: Remove ventilator from service.
Alarm 10. Controller Failure
Alarm message ....................Controller Failure
Alarm sub message..............Controller has reset
unexpectedly
Priority of alarm ....................10
Monitor mode .......................All
Can alarm be muted ............No
Sounder Priority....................High
Alarm Description:
If a command could not be sent to the alarm
generator this message is produced.
Action: Remove patient to alternative form of
ventilation, then remove ventilator from
service.