EasyManua.ls Logo

SLE 2000 - Pre-setting Alarm Limits; Power On Automatic Test; Pressure Display Verification

SLE 2000
66 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 20 of 66
8.2 Pre-setting alarm limits
Set the CPAP alarm control to -5 cmH
2
O and the PIP alarm control to 60 cmH
2
O
8.3 Power On Automatic Test
Set ventilator mode switch to CPAP.
The ventilator will now carry out a self test as follows.
Power LED shows green and an automatic test of functions & alarms is initialised.
Firstly the displays and audible alarms are turned on for approximately two seconds to
demonstrate that they are operable, then for a further three seconds all the digital
displays show a sequence of numbers from O to 9 before returning to their normal
state.
Note: Block LED does not come on during this sequence.
8.4 Pressure Display
Set CPAP/PEEP regulator to give 10 cmH
2
O on the pressure
display with the pressure display selector switch in the MIN
position.
Set the INSP pressure to 40 cmH
2
O
Set the INSP Time to 0.5min
Ensure that :-
LED Pressure display reads 10 cmH
2
O ± 2 cmH
2
O
CPAP/PEEP gauge reads 10 cmH
2
O
Inspiratory cycle is initiated when manual breath button is pressed. (The pressure
display will read 50 cmH
2
O and then return to 10cmH
2
O)
Press reset to cancel the cycle fail alarm.

Related product manuals