20. Technical Specification
20.1 Conventional Ventilation
Modes: CPAP,CMV,PTV,SIMV
BPM Ranges: 1-125 or 126 - 250 breaths per minute (1BPM
STEPS) selected via rear security key switch
Inspiratory Time: 0.1-3.0 or 0.01-0.3 seconds
I:E Range : 9.9:1 - 1:99 calculated from BPM and INSP.
TIME settings.
CPAP Pressure: 0 cmH
2
O to 15 cmH
2
O minimum.
Inspiratory Pressure: 0 cmH
2
O to 60 cmH
2
O
switched fast or slow rise waveforms.
20.2 Displays
Proximal Airway Pressure Gauge Gauge range -6 to +60 cmH
2
O
Seven segment LED’s: showing BPM, INSP. TIME, I:E RATIO, FIO
2
and PRESSURE (max., mean or min.).
Indicator LEDs: POWER: green LED indicates power on.
SYSTEM FAIL: Indicates main processor
system fail.
TRIGGER BACK-UP: Indicates a machine
delivered breath if patient fails to trigger
ventilator during back-up time window.
MAX., MEAN, MIN.: indicates which value is
being displayed on the 7 segment pressure
display.
MUTE: Indicates that the mute function is
active.
Alarm LED’s LOW CPAP : warning of the pressure dropping
below the LOW CPAP alarm limit
CYCLE FAIL: warning that no breath has been
detected. Ventilator breath is below alarm
threshold.
HIGH: warning of the PIP alarm limit being
exceeded
BLOCK: warning of a block in the fresh gas
supply limb of patient circuit.
LEAK : warning of a leak in the fresh gas
supply limb of the patient circuit.
Driving Pressure Gauge:
(Scaled for resultant inspiratory
pressures above PEEP)
Gauge range :0-60 cmH
2
O plus PEEP level