Page 42 of 66
11. Auxiliary Output
Warning No external voltage should be applied to the auxiliary socket. Any
connections to this socket must be approved by SLE and screened to comply
with EMC regulations. Ensure protection cap is fitted when socket is not in use.
Analogue signal outputs are available at the rear of the electronics module, on the 7
pin DIN connector Auxiliary O/P.
Buffered airway pressure. Pin 1 1 - 7 volts corresponding to 0 to 60 cmH
2
O
FIO
2.
Pin 2 0 - 4 volts corresponding to 0 to 100%
Common ground. Pin 3
All other pins must be left unconnected. SLE reserve the right to use these pins for
other uses.
• Only SLE approved equipment may be connected to this auxiliary output socket.