20.3 Controls
Ventilation mode switch: OFF, ALARM TEST/CPAP, CMV, PTV, SIMV
BPM Control (ten turn): either 1-125BPM or 126-250BPM
INSP. TIME Control (ten turn): either 0.10-3.00 seconds (min exp time: 0.25
seconds) or 0.01-0.3 seconds (min exp time:
0.12)
Pressure Display Switch: MAX., MEAN, MIN.
Pressure Wave Switch: Slow or Fast rise
Manual Breath Pushbutton
Alarm Mute and Reset
Pushbuttons:
Mute active for 60 seconds (approx.)
Trigger Sensitivity Control: Range: 2ml/0.5 sec max. To 10ml/0.5sec min.
Using SLE N2188 patient circuit.
Pressure alarm setting controls: LOW CPAP: -5 to 15cmH
2
O
PIP :10 to 60 cmH
2
O
Pressure controls: INSPIRATORY
CPAP/PEEP
Air/Oxygen Blender Control 21-100% ± 3%