8.8 Alarm Test
8.8.1 Low CPAP
Set the following ventilator conditions :-
Return the mode switch to CPAP.
Set the CPAP/PEEP regulator to 5 cmH
2
O.
Increase the Low CPAP alarm control to 10 cmH
2
O. The Low
CPAP alarm will sound and the LED in the alarm panel will
illumuniate.
Reset the Low CPAP alarm control below 5 cmH
2
O. The
Low CPAP audible alarm will cancel but the LED in the alarm
panel will remain illumuniated.
Press the reset button to cancel the visual alarm.
8.8.2 High and Cycle Fail Alarm
Set the PIP alarm control to 30 cmH
2
O.
Set the Pressure display switch to PEAK
Advance the mode switch to CMV
The ventilator will start to cycle
Adjust the Inspiratory pressure regulator to give a pressure of 30 cmH
2
O on the
pressure LED display.
The ventilator is ready for the High and Cycle Fail Alarm tests
Decrease the PIP alarm control to 20 cmH
2
O. This will trigger the HIGH visual and
audible alarm and the LEAK visual alarm
Increase the PIP alarm control to 30 cmH
2
O. The HIGH visual and audible alarm
and the LEAK visual alarm will continue
Press the reset button to cancel the alarms.
Note: The Fresh Gas flow is reduced from 5LPM to1LPM in
the High Alarm condition.