(Model K) Page 137
18.1.18 Functional Test of SIMV Mode
Step 57 For the functional test set the following parameters in the SIMV preview mode:
BPM..................... 20
Ti.......................... 1 sec
PEEP ................... 4 mbar
PIP....................... 25 mbar
FiO
2
..................... 21%
After pressing the confirm button the user will be presented with the
SIMV screen.
Check that:
PSV setup........... is set to OFF
TTV...................... is set to OFF (This button will not appear
when functional testing with no flow sensor fitted).