EasyManua.ls Logo

SLE SLE4000 - Appendix 2. Check list

SLE SLE4000
312 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
(Model K) Page 281
Appendix 2. Check list
The ventilator must pass all the test’s.
Pass Fail
Step 1: Complete power fails alarm test.
Audible alarm triggered and cancelled.
Step 2: Setting FiO
2
.
Measured value corresponds to set.
Step 3: Test of Touch Screen Beep
Touch screen beeps.
Step 6: Oxygen alarm test.
High and Low alarms triggered
Step 11: Function and alarm testing.
Test lung inflates
Step 12:Function and alarm testing.
Analyser corresponds to set pressures
Step 13: Function and alarm testing.
Apnea alarm triggered after 20 seconds
Step 14: Function and alarm testing.
Manual breath delivered
Step 15: Function and alarm testing.
Apnoea alarm triggered after 10 seconds
Step 17: Function and alarm testing.
Continuing positive pressure alarm
is triggered after 10 seconds.
Step 21: Function and alarm testing.
A waveform appears in both waveform
windows
Step 22: Function and alarm testing.
Analyser alternates between set pressures
Step 23: High pressure alarm.
Alarm triggered and cancelled

Table of Contents

Related product manuals