Page 282 (Model K)
Pass Fail
Step 24: Low Pressure alarm.
Alarm triggered and cancelled
Step 25: Breath Not Detected Alarm.
Alarm triggered and cancelled
Step 26: Leak alarm.
Alarm triggered and cancelled
Step 27: Block alarm.
Alarm triggered and cancelled
Step 28: Mains failure alarm.
Alarm triggered and cancelled
Step 29: Gas supply alarms.
No Air supply Alarm triggered and cancelled
Step 30: Gas supply alarms.
No O
2
supply Alarm triggered
Step 31: Gas supply alarms.
No Gas Alarm triggered and cancelled
Step 32: Flow sensor disconnect alarm.
Alarm triggered and cancelled
Step 33: Cycle fail alarm.
Alarm triggered and cancelled
Step 38: CMV functional test.
ventilator is cycling, waveforms
appear and I:E ratio = 1:1
Step 40: Wave shape modified.
Pressure wave modified