Page 276 (Model K)
Item 12: Pressure Triggering Verification
Item 13: Gas Fail Detection Verification
Item 14: Blender and Oxygen Monitoring Verification
CPAP
Item 15: Battery Charge Verification
Item 16: Battery Life Verification
Results of battery life test:
Pressure Trigger Verified:
Gas Fail Detection Verified:
Set Oxygen concentration Lower Limit Actual
Reading
Upper limit
21% 20% 24%
30% 27% 33%
40% 37% 43%
50% 47% 53%
60% 57% 63%
70% 67% 73%
80% 77% 83%
90% 87% 93%
100% 97% 100%
Lower limit Actual reading Upper limit
13.7V >13.7V
Start Time:
Battery Low Alarm Time:
Power Fail Time: