(Model K) Page 277
Item 17: Total Power Fail Alarm Test
Item 18: Pressure Calibration Verification
a) CPAP
b) CMV
Total power fail alarm test: Initial
PIP Setting Lower limit Actual reading Upper limit
0 mbar 0 mbar 1 mbar
5 mbar 4 mbar 6 mbar
10 mbar 9 mbar 11 mbar
15 mbar 14 mbar 16 mbar
20 mbar 19 mbar 21 mbar
PIP Setting Lower limit Actual reading Upper limit
10 mbar 9 mbar 11 mbar
20 mbar 19 mbar 21 mbar
30 mbar 29 mbar 31 mbar
40 mbar 39 mbar 41 mbar
50 mbar 49 mbar 51 mbar
60 mbar 59 mbar 61 mbar