14
Pump Setup
Always check the battery compartment for uid
or debris before inserting the batteries and do
not allow any uid or debris to fall into the
battery compartment. Fluid or debris in the
battery compartment may damage the battery
contacts and could result in loss of power and
nondeliveryofdrugand,dependingonthetype
of drug being administered, could result in
death or serious injury to the patient.
If the pump is dropped or hit, the battery
door may become broken or damaged. Do not
use the pump if the battery door is damaged
because the batteries will not be properly
secured; this may result in loss of power,
nondeliveryofdrugand,dependingonthetype
of drug being administered, could result in
death or serious injury to the patient.
Toinstallthebatteries:
1. Make sure the
pump is stopped
or powered off.
Using your fingers,
the pump key, or a
coin, turn the knob
on the battery door
counter-clockwise and open the battery door.
2. Hold the pump at an
angle and place 4 AA
batteries in the pump,
from the bottom up
(see picture). Match
the + and - markings
on the new batteries
with the markings on
the pump.
OR: If using a
rechargeable battery
pack, insert it into the
pump as shown.
Pump Setup
Installing the Batteries
AA 1.5 volt primary (non-rechargeable) alkaline
batteries or the Smiths Medical rechargeable battery
pack are recommended for use in the CADD
®
-Solis
pump.
NOTE:Smiths Medical does not recommend
mixing new and used batteries; doing so may affect
low battery alarm times. Always select four new
batteries when replacing them.
CAUTION:InspecttheAAbatteriesfordamage
or wear to the metal or plastic insulation prior to
use, or aer the pump has been dropped or hit.
Replace the batteries if any damage is noted.
The pump retains all programmed values while the
batteries are removed. The pumps batteries must be
in place during delivery. If the batteries are removed
while the pump is delivering, and an AC adapter is
connected, delivery will stop. If an AC adapter is not
connected and the batteries are removed, delivery
will stop and the pump will lose power.
Dispose of used batteries in an environmentally safe
manner, and according to any regulations which
may apply.
WARNING:
Do not use rechargeable NiCd or nickel metal
hydride (NiMH) batteries. Do not use carbon
zinc(“heavyduty”)batteries.eydonot
providesucientpowerforthepumpto
operate properly, which could result in death or
serious injury to the patient.
Alwayshavenewbatteriesavailablefor
replacement.Ifpowerislost,nondeliveryof
drug will occur and, depending on the type of
drug being administered, could result in death
or serious injury to the patient.
ere is no pump alarm to alert users that a
battery has not been properly installed. An
improperly installed battery could result
inlossofpowerandnondeliveryofdrug
and, depending on the type of drug being
administered, could result in death or serious
injury to the patient.

Table of Contents

Other manuals for Smiths Medical CADD-Solis

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Smiths Medical CADD-Solis and is the answer not in the manual?

Smiths Medical CADD-Solis Specifications

General IconGeneral
BrandSmiths Medical
ModelCADD-Solis
CategoryMedical Equipment
LanguageEnglish

Related product manuals