36
References and Troubleshooting
Alarms and Messages, Alphabetical List
Alarms and Messages Alarm Priority Description / Corrective Action
(Screen is blank and alarm is
sounding)
High The pump was delivering and the batteries were removed or the
battery door was opened. The pump has lost power and is no
longer delivering. Clear this alarm by turning the pump back
on, or the alarm will stop after the power has been off for a
minimum of 2 minutes.
A setting was edited, but not
saved, and the edit was lost.
Medium A parameter was being manually edited, but it wasnt saved and
the pump reverted to the home screen. Select acknowledge to
clear the alarm, and if appropriate, edit the parameter and save.
AC Adapter disconnected. Low The AC adapter was disconnected and the pump is being
powered by the 4 AA batteries or the rechargeable battery
pack. Select acknowledge to clear the alarm or the alarm will
automatically clear after 5 seconds. If desired, reconnect the AC
adapter.
Air in-line detected. Press
Acknowledge’ then prime
tubing.
High The air detector has detected air in the fluid path; the fluid
path may contain air bubbles. The pump was delivering and is
now stopped and will not run. Select acknowledge to clear the
alarm, then:
• Ifthefluidpathcontainsairbubbles,closetheclampsand
disconnect the fluid path from the patient. Then follow the
instructions for removing air by priming (see Prime Tubing on
page 24 for more information on priming). Restart the pump.
Battery depleted. Pump
stopped.
High The pump was delivering and is now stopped and the battery
power is too low to operate the pump. If the AC adapter is
attached, select acknowledge to clear the alarm.
• Removethebatteries.
• Install4newAAbatteriesorafullychargedrechargeable
battery pack. In order to start delivery, good batteries must
always be installed, even when an external source of power is
connected.
• Ifappropriate,restartthepump.
Battery depleted. Pump will not
run.
Medium The battery power is too low to operate the pump. If the AC
adapter is attached, select acknowledge to clear the alarm.
• Removethebatteries.
• Installafullychargedrechargeablebatterypackor4new
AA batteries. In order to start delivery, good batteries must
always be installed, even when an external source of power is
connected.
Battery low. Replace battery. Low The rechargeable battery pack or the 4 AA batteries are low
but the pump is still operable. Select acknowledge to clear the
alarm or the alarm will automatically clear after 5 seconds.
• Rechargeorchangetherechargeablebatterypackorreplace
the 4 AA batteries soon.

Table of Contents

Other manuals for Smiths Medical CADD-Solis

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Smiths Medical CADD-Solis and is the answer not in the manual?

Smiths Medical CADD-Solis Specifications

General IconGeneral
BrandSmiths Medical
ModelCADD-Solis
CategoryMedical Equipment
LanguageEnglish

Related product manuals