References and Troubleshooting
Alarms and Messages Alarm Priority Description / Corrective Action
Battery removed. Pump
stopped.
High The rechargeable battery pack or the 4 AA batteries were
removed. The pump was delivering and is now stopped. Select
acknowledge to clear the alarm.
• Installafullychargedrechargeablebatterypackor4new
AA batteries. In order to start delivery, good batteries must
always be installed, even when an external source of power is
connected.
• Ifappropriate,restartthepump.
Battery removed. Pump will
not run.
Medium The pump is stopped and the rechargeable battery pack or the
4 AA batteries were removed, but the pump is still powered by
the AC adapter. Select acknowledge to clear the alarm.
• Installafullychargedrechargeablebatterypackor4new
AA batteries. In order to start delivery, good batteries must
always be installed, even when an external source of power is
connected.
Cannot start pump with a
depleted battery.
Medium The battery power is too low to operate the pump. In order
to start delivery, good batteries must always be installed
even when an external source of power is connected. Select
acknowledge to clear the alarm. Install 4 new AA batteries or a
fully charged rechargeable battery pack. If appropriate, start the
pump.
Cannot start pump with a
reservoir volume of zero.
Medium The reservoir volume in the pump is set to zero. Select
acknowledge to clear the alarm. Edit the reservoir volume to
the correct value. If appropriate, start the pump.
See Patient Specific Parameters on page 29 for more information
about the reservoir volume.
Cannot start pump with
rechargeable battery reached
end of use.
Medium The rechargeable battery pack is at the end of its life. It has been
discharged and recharged so many times that it is no longer
able to hold a good charge. Remove the battery from service.
In order to start delivery, good batteries must always be
installed even when an external source of power is connected.
Select acknowledge to clear the alarm. Install 4 new AA
batteries or a new rechargeable battery pack. If appropriate,
start the pump.
Cannot start pump without a
battery.
Medium The pump does not have any batteries installed. In order
to start delivery, good batteries must always be installed
even when an external source of power is connected. Select
acknowledge to clear the alarm. Install 4 new AA batteries or a
rechargeable battery pack. If appropriate, start the pump.
Cannot start the pump with air
in-line. Prime tubing.
Medium The air detector has detected air in the fluid path directly
under the air detector; the fluid path may contain air bubbles.
Select acknowledge to clear the alarm, then:
• Ifthefluidpathcontainsairbubbles,closetheclampsand
disconnect the fluid path from the patient. Then follow the
instructions for removing air by priming (see page 24 for more
information).
• Ifappropriate,startthepump.

Table of Contents

Other manuals for Smiths Medical CADD-Solis

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Smiths Medical CADD-Solis and is the answer not in the manual?

Smiths Medical CADD-Solis Specifications

General IconGeneral
BrandSmiths Medical
ModelCADD-Solis
CategoryMedical Equipment
LanguageEnglish

Related product manuals