52
References and Troubleshooting
Default Factory Settings
The first time you use the pump, the protocol will
be set to the factory default. If you desire to reset the
pump to the factory default, you may do so at any
time (refer to the Administrator Settings Guide for
more information on how to reset the pump to the
factory default).
The following chart details the factory default
parameters:
Programming Units mL
Continuous Rate 0 mL/hr
PCA Dose Amount 0 mL
PCA Dose Lockout
Period
1 hour
Delivery Limit Method Not in Use
Reservoir Volume 1 mL
Reservoir Volume Reset
Value
100 mL
Reservoir Volume Low
trip point
5.0 mL
Reservoir Volume Low
Alert Type
Insistent and One Time
Only
Repeat Reservoir Volume
Empty Alert Type
One Time Only
Clinician Bolus Amount 0 mL
Maximum Fluid Delivery
Rate (combined bolus
and continuous)
175 mL/hr
Pump Stopped Alert Type Informational
Upstream Sensor On
Downstream Occlusion
Sensitivity
Low
Air Detector Enabled Yes
Air Detector Sensitivity Low
Date Format mm/dd/yy
Preventative
Maintenance Reminder
Off
Sound Volume High
Unlock Security Code ** Text Omitted **
Clinician Security Code ** Text Omitted **
Administrator Security
Code
See your CADD
®
-Solis
system administrator
for default security
codes
Keypad Security Code only
Backlight Intensity 3
Color Theme Black
Sound Theme Standard
Key Beeps Enabled Yes
Manual Mode
Programming Code
Administrator Code