49
References and Troubleshooting
MaximumDeliveryRate(combinedbolusand
continuous)
40 to 175 mL/hr in increments of 1 mL.
Key Beeps
On or off.
ResVolLowTripPoint
1 to 999 mL in increments of 1 mL.
ResVolEmptyAlarm
Insistent and one time only or non-insistent
and repeating.
Pump Stopped Alarm
Informational or high priority.
Air Detector
Turned on or turned off.
AirDetectorSensitivity
Low Sensitivity - Single bubble greater than
400μL .
High Sensitivity - Single bubble greater than
150μL .
Alarm Volume
High, medium, or low.
PM(PreventativeMaintenance)Reminder
Interval: 1 to 24 months in 1 month increments.
Enable: On or off.
Custom Keypad Code
001 to 899 in increments of 1.
Custom Clinician Code
001 to 899 in increments of 1.
Custom Administrator Code
001 to 899 in increments of 1.
Date Format
US standard (mm/dd/yy) or European standard
(dd/mm/yy).
Time
00:00 to 23:59.
DownstreamOcclusionSensitivity
High Sensitivity - once the high pressure alarm
threshold is reached, the downstream occlusion
alarm is triggered immediately.
Low Sensitivity - once the high pressure alarm
threshold is reached, the downstream occlusion
alarm is delayed for 2 seconds, this allows for
the pressure to settle before a possible alarm
(if the pressure settles below the high pressure
alarm threshold before the 2 second delay is
complete, the alarm will not happen).
Upstream Occlusion Sensor
On or off.
NOTE: The upstream occlusion sensor
is automatically disabled during use with
medication cassette reservoirs.
Electromagnetic Emissions and Immunity
Declarations
Guidance and manufacturer’s declaration - electromagnetic emissions
The CADD®-Solis pump is intended for use in the electromagnetic environment specified
below. The customer or the user of the CADD®-Solis pump should assure that it is used in such
an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1 The Pump uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B The Pump is suitable for use in all establishments,
including domestic establishments and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not applicable
Compliance using:
1. 100-240 VAC 50/60Hz to 7VDC switching
desktop AC adapter, with an AC power cord
length of 1.8 m (6 feet).
2. Rechargeable battery pack.
3. Remote dose cord with a length of 152 cm ± 5
cm (60 in. ± 2 in.).
4. USB cable length of less than 2 m (6.5 feet).
WARNING:
e use of power supplies and a remote
dose cord other than those listed in the
electromagnetic emissions declaration may
result in increased emissions or decreased
immunity of the pump.
e pump should not be used adjacent to or
stacked with other equipment. If adjacent
or stacked use is necessary, the user should
verifynormaloperationofthepumpinthe
conguration in which it is to be used.

Table of Contents

Other manuals for Smiths Medical CADD-Solis

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Smiths Medical CADD-Solis and is the answer not in the manual?

Smiths Medical CADD-Solis Specifications

General IconGeneral
BrandSmiths Medical
ModelCADD-Solis
CategoryMedical Equipment
LanguageEnglish

Related product manuals