MIC-5010, MIC-5005 USER MANUAL
17
- 5 kV: 1 kV, 2 kV, 3 kV, 4 kV and 5 kV.
The end result for each of the 5 measurements is saved which is signalled by a beep and an
appropriate icon.
Set the rotary switch of function se-
lection in the SV position. The me-
ter is in the voltage measurement
mode.
By pressing the MENU button the maximum
measurement voltage selection, the duration for
individual measurements and the I
ISO
current may
be selected.
The buttons and
are used to set the pa-
rameter value, while the and buttons move
to the next parameter.
The setting sequence is as follows:
maximum (final) measurement voltage: 1 kV, 2.5 kV and 5 kV,
duration of individual measurements in the range between 30 s...5
min,