47Instruction for use | Supercal 5
9.3 Error Messages
If several errors are present at the same time, the individual error mes-
sages are added together and displayed.
If an error occurs for more than one hour, it is stored in the error memory
with date and time (start of error) and duration (in minutes). If an error
is present for less than 60 minutes, it is deleted automatically without
being saved.
Code Error LCD Reference Description Procedure Troubleshooting
1 A/D reference 1 Temperature reference 1 A/D:
A temperature sensor cable is
interrupted or not connected.
Check if the sensor probe is connected.
If yes, the sensor probe is defective.
2 A/D reference 2 Temperature reference 2 A/D:
A cable of the temperature sensor
is interrupted or not connected.
Check if the sensor probe is connected.
If yes, the sensor probe is defective.
4 Temperature sensor 1
A/D error
Temperature reference 1 A/D:
A cable of the temperature sensor
is connected but its value can not
be read out.
Swap the probes. If Err2 appears afterwards,
probe 1 is defective or the wrong type.
8 Temperature sensor 2
A/D error
Temperature reference 2 A/D:
A cable of the temperature sensor
is connected but its value can not
be read out.
Swap the probes. If Err1 appears afterwards,
probe 2 is defective or the wrong type.
16 Range Sensor 1 Temperature sensor 1 < = min.
Range error
Check the actual temperature being meas-
ured and the physical installation of the
sensor.
32 Range Sensor 1 Temperature sensor 1 > = max.
Range error
Check the actual temperature being meas-
ured and the physical installation of the
sensor.
64 Range Sensor 2 Temperature sensor 2 < = min.
Range error
Check the actual temperature being meas-
ured and the physical installation of the
sensor.
128 Range Sensor 2 Temperature sensor 2 > = max.
Range error
Check the actual temperature being meas-
ured and the physical installation of the
sensor.
512 Flow Saturation
1024 OPEN CASE The SC5 is open Close the case
2048 Main power cut Power outage Check the main power supply connection.
M1 Power Supply Error in module 1:
module error
M1 Unsupported
Slot left error
M2 Power Supply Error in module 2:
module error
M2 Unsupported
Slot right error
9.3.1 Overview of Error Messages
Note: Any combination of errors corresponds to the arithmetic sum of each code. The error
code is displayed on the electronic device’s display.