EasyManuals Logo

Sontex Supercal 5 I User Manual

Sontex Supercal 5 I
68 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #49 background imageLoading...
Page #49 background image
Instruction for use | Supercal 5
11. Technical Information
11.1 Calculator Supercal 5
Temperature Measurement
Pt500 according to EN 60751
2- or 4-wire
Absolute temperature range
Approved range
Homologation range
Response limit
Temperature resolution t

Environment class A
 °C to 200 °C
1 °C to 200 °C
3 K to 150 K
0,2 K
0,1 K
0.01 K
E1/M1
Temperature Measuring Cycle
Battery operated
Mains operated
10 s
3 s
Medium Temperature
Operation
Storage and transport
5 °C to 55 °C

Display
Illuminated dot-matrix
128 × 64 pixels
Display Units
Energy
Volume
Additional pulse inputs
Temperature
kWh, MWh, MJ, GJ, kBtu, MBtu, Mcal, Gcal
L, m
3
, gal (US), kgal (US), ft
3
(US)
Energy or volume
°C, °F
Lifespan Supply Modules
w/o supply
D battery


Degree of Protection
IP Code

Pulse Inputs
F
requencies
Without supply
D battery
External Mains
Input voltage
maximum 5 Hz
maximum 200 Hz
maximum 200 Hz
0 to 30 V
Pulse Outputs
Frequencies
Without supply
D battery
External Mains
Output voltage
maximum 5 Hz
maximum 200 Hz
maximum 200 Hz
0 to 60 V
Optical interface
Interface
according to IEC 62056-21:2002
NFC Interface
Interface
according to ISO/IEC 14443 Type A
M-Bus interface
Interface
Baud rate
Galvanic isolation
according to EN 13757-2/3

3.75 kV
Technical Data

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sontex Supercal 5 I and is the answer not in the manual?

Sontex Supercal 5 I Specifications

General IconGeneral
Power supply3.6 V Lithium battery
Measuring principleUltrasonic
MID approvalYes
Communication interfaces (optional)M-Bus, LoRaWAN
Measurement AccuracyClass 2 (EN 1434)
Pipe Diameter RangeDN15 to DN40
OutputsPulse output
Nominal flow rate qp1.5 m³/h