Important EMC notices for use in the medical environments
The MCC-3000MT needs special precautions regarding EMC and needs to be installed and
put into service according to the EMC information provided in this instructions for use.
The portable and mobile RF communications equipment such as cellular phones can affect the
MCC-3000MT.
Warning
The use of accessories and cables other than those specified, with the exception of replacement
parts sold by Sony Corporation, may result in increased emissions or decreased immunity of the
MCC-3000MT.
List of cables used for EMC tests
Type of cable Specifications
DC power cable 1.5 m, non-shielded
CCMC-T15 camera cable 15 m, shielded
BNC Cable (5C-FW) 1.5 m, shielded
D-Sub 9P Cable (RS232C) 1.5 m, shielded
Guidance and manufacturer’s declaration - electromagnetic emissions
The
MCC-3000MT is intended for use in the electromagnetic environment specified below.
The customer or the user of the MCC-3000MT should assure that it is used in such an
environment.
Emission test Compliance Electromagnetic environment-guidance
RF emissions
CISPR 11
Group 1 The
MCC-3000MT uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B The
MCC-3000MT is suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies