achieve the working temperature, in order to obtain the accurate resultes.
2.4 Calibration
In order to obtain the accurate resultes, it is need to calibration according to the local acceleration of gravity.
Necessary situations to calibrate the balance:
·Before the first measurement
·After weighing for a period
·Weighing location changed
Should start the balance for 60 minutes before calibration to achieve the working temperature, in order to obtain
the accurate.
External Calibrtaion
Clear weighting pan, perpare calibration weights.
Press , untill display " ".
The needed calibraion weight value will flicker on the display
screen.
Put the right calibration weight on the scale pan. When it
shows , remove weight.
When it displays " " and emits prompting
voice, the calibration is over. The balance return to the
working situtaion. When shows “ , the balance is
ready.
Hint
Press short time, could interrupt calibration
any time.