EasyManua.ls Logo

Soptop FA2204 - 5.7 Recover Factory Setting; Technical Parameters, Function Options, Parts, Dimension

Default Icon
37 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
26
”flickering means that there is nothing on the weighing pan. Press shortly. When send
out voice ofdu” and stop flickering, it is means finishing corretion. Then the balance auto turn to the next
correcting point “ ” , it need to add 50g weight. Then press shortly to ensure. After
finishing all the correction point, the balance will display . Now the balance is still processing the
data, please wait. Finishing data processing, the balance will shut down automatically.
5.7 Recover Factory Setting
If customer need to recovery the factory setting, please enter the balance setting mode when starting. Choose
C313 to recover, then the balance will shut down automatically. Also it need to do linear correction.
6 TECHNICAL PARAMETERS, FUNCTION OPTIONSL,
PARTS OPTIONAL, DIMENSION
6.1 Technical Parameters
Standard Configuration Protection Class
·Power100-240VAC50/60Hz ·dustproof, waterproof
OutputDC12V600mA ·Pollution prevention level: 2
·Installation category: 11
Raw Material Environment Requirement:
·Basspack alloy, painting ·Environment Conditions
·Coverplastic (ABS/PC)
·Environment Temperature 20℃±2.5 fluctuations in
temperature is not more than 1/h
·Weighing panstainless steel 20℃±7.5℃,fluctuations in temperature is not more
than 5/h
·Relatice Humidity 50%~75% 40%~80%
·Working Voltage 124DC 12VDC
TARE
/