24
5.2 Balance Function Setting List(* measn factory set)
Page Line Word Function Optional
C 1 1 1
Adapting the balance to ambient
conditions
Very stable
C 1 1 2 stable *
C 1 1 3 Unstable
C 1 1 4 Very unstable
C 1 2 1
Ambient conditions
Stability range
0.25 digit
C 1 2 2 0.5 digit
C 1 2 3 1 digit *
C 1 2 4 2 digit
C 1 2 5 4 digit
C 1 2 6 8 digit
C 1 2 7 16 digit
C 1 2 8 32 digit
C 1 2 9 64 digit
C 1 3 1
Display
Display all decimal places *
C 1 3 2 Not displaying the last decimal place
C 1 3 3 Display the last decimal place stability
C 1 3 4 Display all decimal places at stability
C 1 4 1
Remove the container
Remove regardless of stability
C 1 4 2 Remove at stability *
C 1 5 1
Auto-zero function
Auto-zero on *
C 1 5 2 Auto-zero off
C 2 1 1
RS232C output
Print on request regardless of stability
C 2 1 2 Print on request after stability *
C 2 1 3 Auto print no stoppable regardless of stability
Auto print not stoppable at stability
C 2 2 1
Baud
1200
C 2 2 2 2400
C 2 2 3 4800
C 2 2 4 9600 *
C 2 2 5 19200