9
Different model balance has different calibration weight, please see parameter list at part six.
Please keep packing materials properly for transport
2.2 Install, level Adjustment
Best placed place
For high resolution analytical and precise balance, the right
placed place is the key for accurately weighing. So please
ensure that:
·Stable and no vibration position, as far as possible be level
·Avoid direct sunlight
·Avoid strenuous temperature variations
·Avoid air convection
Best placed place: the corner of shelter, stable table, as far as
possible from the door, window, radiator and the outlet of air
condition
Through regulating the leveling foot, make the level bubble moving to the
center of circle. Please adjust balance in the stable table, avoid the
position of the large air flow, virbration and high rate temperature change.
(as shown)
Notice: Please adjust level, when the balance move to a new place.
2.3 Safety Guidelines/Power
• Do not use the balance equipped with standard AC power in the
damage area.
• Please confirm the power adapter whether is match to your local
volatge before getting through the power.
• The balance can only be used in dry enviroment.
Power connection:
→Plug jack-plug of AC adapter in the socket of balance, another side
connect the power.
→Turn on the power, a buzzer emits beep. The balance could be used.
Notice:The balance should be warmed-up for 60 minutes in first time, to