91369 Service Manual 2-31
Setup
User Access
The USER ACCESS key allows the system administrator to preset certain functions and features of the
monitor for availability to non-privileged-access users.
• PATIENT TYPE / ON/OFF — Enables (ON) or disables (OFF) the “Patient Type” selection in the Admit/
Discharge dialog box.
• PARAMETER CONFIG / ON/OFF (bedside monitors only) — Displays (ON) or removes (OFF) the
PARAMETER CONFIG key in the Monitor Config menu.
• RECORDING DURATION / ON/OFF — Displays (ON) or removes (OFF) the RECORDING DURATION
key in the Recorder Config menu.
• SUBNET ACCESS / ON/OFF — Select ON to display keys for other care areas (subnets) within bed
selection windows for features such as Alarm Watch, Remote View, or Screen Format.
• DEFAULT ENG. SAV MODE / ON/OFF — Controls the monitor’s default-energy-saving power mode ON
or OFF and changes the setting of the ENERGY SAVING MODE / ON/OFF key that appears when the
monitor is operating on battery power.
Alarm Setup
Touch the ALARM SETUP key to display the following keys:
• REMOTE ACCESS / ON/OFF — Select ON to allow alarm limits from this monitor’s parameters to be
changed remotely (from central monitors or from bedside monitors via Remote View).
Selecting OFF ensures that alarm limits for this monitor’s parameters can only be changed at this monitor.
• ALARM SUSPEND / ON/OFF — Select ON to enable access to the TONE RESET/ALM SUSPEND key’s
Alarm Suspend function.
• TREND SUSPEND / ON/OFF — Select ON to allow trending to occur when all alarms are suspended via
the ALARM SUSPEND key.
• ALARM RELAY — Allows characteristics of the monitor’s external alarm relay to be defined. When an
alarm occurs, this relay can activate an external device to identify which monitor is in alarm. In general
terms, monitors activate their alarm relay whenever an alarm is occurring on that monitor.
- RELAY TIMEOUT / 0 SEC/10 SEC — Select 0 SEC to deactivate the alarm relay when the alarm ends.
Select 10
SEC to deactivate the alarm relay 10 seconds after the alarm ends.
- FLASHING/STEADY ON — Controls whether the alarm relay is intermittently activated (FLASHING) or
is continuously activated (STEADY ON) when an alarm occurs. Intermittent activation is normally used
when connecting to an external light. Continuous activation is normally used when connecting to a
nurse call button (or similar equipment).
Note:
Setting the monitor ALARM RELAY to either FLASHING or STEADY ON does not affect the 91369
embedded nurse alert light.
- ALARM LEVEL — The alarm relay can be activated for all alarms or for alarms at or above the selected
priority only. Selections of HIGH, MEDIUM, and LOW are available. For example, selecting MEDIUM
results in activation of the alarm relay for HIGH and MEDIUM priority alarms, but not for LOW priority
alarms.