Preparation
Instructions for use • S-PILOT
®
• PYB194_EN_V1.0_02-2021_IFU_CE-MDR 22
5.6.2
Connecting the product as a stand-alone solution
1. Attach the SCB connecting cable (20090170) to AUTOCON
®
II 400 and to a SCB
connection on S-PILOT
®
.
2. Attach connecting cable (UP004) to AUTOCON
®
III 400 or ConMed System 2450 or 5000
and to the remote control connection on S-PILOT
®
.
3. Attach connecting cable (UP005) to COVIDIEN Force Triad or Valleylab Force FX and to
the remote control connection of S-PILOT
®
.
5.7 Connecting the product for suctioning fluid
1. Connect the control cable (27610070) to UNIDRIVE
®
S III and to the footswitch socket on
S-PILOT
®
.