Electromagnetic compatibility
Instructions for use • S-PILOT
®
• PYB194_EN_V1.0_02-2021_IFU_CE-MDR 32
9 Electromagnetic compatibility
9.1 General notes on the operating environment
The product is suitable for use in professional healthcare settings. Professional healthcare
facilities include physician offices, dental offices, limited care facilities, freestanding surgical
centers, freestanding birth centers, multiple treatment facilities, hospitals (emergency rooms,
patient rooms, intensive care, surgical rooms, outside the HF-shielded room of an ME system
for MRT).
The emission characteristics of this product make it suitable for use in professional
healthcare facilities as well as in a residential environment (CISPR 11 Class B). This
product offers adequate protection to radio communication service. In the rare event of
interference to the radio transmission operation, the user might need to take mitigation
measures, such as relocating or re-orienting the product.
9.2 Accessories and cables
Accessories and cables for EMC compliance
Type Shield Length [m] Ferrite Use
PA No >3 No Potential equal-
ization
SCB cable Yes ≥ 1 No Connection to
the SCB net-
work, other SCB
devices
Power cord No 3 No Mains connec-
tion
20014130 Yes 3 No One-pedal
footswitch
27610070 Yes 0.42 No Control cable
(connecting ca-
ble between
S‑PILOT
®
and
UNIDRIVE
®
S III)
20701070 Yes 1 No Control cable
(connecting ca-
ble between
S‑PILOT
®
and
UNIDRIVE
®
S III /
CALCUSON)
UP004 No 3 No Connecting ca-
ble between S-
PILOT
®
and AU-
TOCON
®
III 400 /
ConMed System
2450 or 5000
UP005 No 3 No Connecting ca-
ble between S-
PILOT
®
and Co-
vidien Force