Faults and messages
Instructions for use • S-PILOT
®
• PYB194_EN_V1.0_02-2021_IFU_CE-MDR 38
10 Faults and messages
10.1 Troubleshooting
Fault Possible causes Actions
Product failed Power supply failure
Have the power supply
checked
Power plug and socket im-
properly connected
Push the power plug
firmly into the socket on
the product
Defective fuse
The fuse must not be
changed by the user:
Contact Service
Inadequate suction power Leakage occurring in tubing
system
Check the tubing line
and replace it if
necessary
Check the sealing cap is
correctly seated
Tube system blocked
Remove the blockage
and replace the tubing
line if necessary
Suction receptacle full
Empty or replace
suction receptacle
Error in self-test System error:
Status LED and buzzer:
T
on
/T
off
1 s/1 s
Contact Service
Motor driver:
Status LED and buzzer:
T
on
/T
off
1 s/3 s
SCB connection lost during
operation, acoustic informa-
tion signal every 10 s
Check SCB connection