Electromagnetic compatibility
Instructions for use • S-PILOT
®
• PYB194_EN_V1.0_02-2021_IFU_CE-MDR 36
Interference immunity
tests
EN/IEC60601 test
level
Compliance
level
Electromagnetic envi-
ronment – guidelines
be less than the compli-
ance level in each fre-
quency range
b
.
d = 1.2√P
80MHz to 800MHz
d = 2.3√P
800MHz to 2.5GHz
Interferences may occur in
the vicinity of devices
marked with the following
symbol:
Note: At 80MHz and 800MHz, the higher frequency range applies.
Note: These guidelines may not apply in all situations. The propagation of electromagnetic
waves is affected by absorptions and reflections of buildings, objects and people.
a
Field strengths from fixed transmitters, e.g. base stations for radio (cellular/cordless) tele-
phones and land mobile radios, amateur radio, AM and FM radio and TV broadcasts cannot
be predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed transmitters, an electromagnetic site survey should be considered. If the measured field
strength at the location where the product is used exceeds the above compliance levels, the
product should be monitored to ensure proper function. If abnormal performance is ob-
served, additional measures may be necessary, such as re-orienting or relocating the prod-
uct.
b
Over the frequency range from 150kHz to 80MHz, field strengths should be less than 3V/
m.
9.6 Table 4 – Emission class and group
Guidelines and manufacturer’s declaration – electromagnetic emissions
The product is intended for use in such an environment as specified below. The customer or
user of the device should ensure that it is used in such an environment.
Interference emission mea-
surements
Conformity Electromagnetic environment–
Guidelines
RF emissions according to
CISPR11
Group1 The product uses RF energy for its in-
ternal function only. Therefore, its RF
emissions are very low and are not
likely to cause any interference affect-
ing nearby electronic equipment.
RF emissions according to
CISPR11
ClassB The product is suitable for use in all
establishments including domestic es-
tablishments and those directly con-
nected to the public low voltage power
supply network that supplies buildings
used for domestic purposes.
Harmonic emissions acc. to
IEC61000-3-2
ClassA
Voltage fluctuations/flicker emis-
sions acc. to IEC 61000-3-3
Compliant