What to do if Stryker SERFAS displays error F7?
- TTraci HintonSep 2, 2025
If your Stryker Medical Equipment displays error F7, it means there is a RAM error. Cycle power to the unit (turn the unit off and then on) to resolve it.
What to do if Stryker SERFAS displays error F7?
If your Stryker Medical Equipment displays error F7, it means there is a RAM error. Cycle power to the unit (turn the unit off and then on) to resolve it.
What to do if Stryker SERFAS displays error F6?
If the Stryker Medical Equipment displays error F6, it means there is a ROM CRC error. You can resolve this by cycling power to the unit (turning it off and then on).
What to do if Stryker SERFAS Medical Equipment displays error F5?
If your Stryker Medical Equipment displays error F5, it indicates a COP Timeout error. To solve it, cycle power to the unit (turn the unit off and then on).
What to do if Stryker SERFAS Medical Equipment displays error F4?
If your Stryker Medical Equipment displays error F4, it indicates a CPU Configuration error. To fix it, cycle power to the unit (turn the unit off and then on).
What to do if Stryker Medical Equipment shows error E8?
If the Stryker Medical Equipment displays error E8, it indicates the RF Current is too high. You should check the probe for damage and replace it if necessary. Also consider replacing the handpiece cable.
What to do if Stryker SERFAS Medical Equipment shows error E7?
If your Stryker Medical Equipment shows error E7, it means the RF Voltage is too high. You should check the probe for damage and replace it if necessary. Also consider replacing the handpiece cable.
What to do if Stryker Medical Equipment displays error F3?
If your Stryker Medical Equipment displays error F3, it indicates an RF generation or measurement error during the power-on test. Try cycling the power to the unit (turn it off and then on).
What to do if Stryker Medical Equipment displays error F2?
If the Stryker Medical Equipment displays error F2, it means that there was a hardware shutdown because current, voltage, or power limits were exceeded. To resolve this, cycle the power to the unit (turn it off and then on).
What to do if Stryker SERFAS displays error F1?
If your Stryker Medical Equipment is displaying error F1, it indicates an RF Current measurement error. To solve this, cycle the power to the unit by turning it off and then on again.
What to do if Stryker SERFAS Medical Equipment displays error E2?
If your Stryker Medical Equipment displays error E2, it means the probe or handpiece cable is not being detected when the footswitch pedal is activated. Try reconnecting the probe or the handpiece cable. If that doesn't work, you might need to replace the probe or the handpiece cable.
Brand | Stryker |
---|---|
Model | SERFAS |
Category | Medical Equipment |
Language | English |
Alerts to potential risks to patient or physician safety if instructions are ignored.
Highlights procedures to avoid damaging the instrument.
Provides supplementary information for easier maintenance or clarity.
Describes the main unit of the SERFAS Energy System.
Explains how to use the CUT function of the system.
Explains how to use the COAG function of the system.
Describes how to adjust the hand control feature.
Instructions for cleaning the generator unit.
Instructions for cleaning the footswitch.
Instructions and warnings regarding probe cleaning.
Guidance on how to dispose of system components.
Lists the technical specifications for the energy probe.
Lists the technical specifications for the energy generator.
Displays graphs showing output power under various conditions.