What to do if Stryker SERFAS displays error F6?
- JJon WilliamsSep 1, 2025
If the Stryker Medical Equipment displays error F6, it means there is a ROM CRC error. You can resolve this by cycling power to the unit (turning it off and then on).

What to do if Stryker SERFAS displays error F6?
If the Stryker Medical Equipment displays error F6, it means there is a ROM CRC error. You can resolve this by cycling power to the unit (turning it off and then on).
What to do if Stryker SERFAS displays error F7?
If your Stryker Medical Equipment displays error F7, it means there is a RAM error. Cycle power to the unit (turn the unit off and then on) to resolve it.
What to do if Stryker SERFAS Medical Equipment displays error F8?
If your Stryker Medical Equipment displays error F8, it means there is an Illegal CPU Instruction. Cycle power to the unit (turn the unit off and then on) to attempt to resolve it.
What to do if Stryker Medical Equipment shows error E3?
If the Stryker Medical Equipment displays error E3, it means multiple switches were pressed simultaneously. Ensure that you press only one switch at a time.
What to do if Stryker SERFAS Medical Equipment displays error F9?
If your Stryker Medical Equipment is showing error F9, it indicates a software error. To resolve this, cycle the power to the unit (turn it off and then on).
What to do if Stryker SERFAS shows error E5?
If the Stryker Medical Equipment displays error E5, and a high impedance is measured, ensure the probe is activated inside saline. If the error persists, replace the probe.
What to do if Stryker Medical Equipment generator temperature is too high?
If the generator temperature is too high, ensure proper airflow on the sides and rear of the generator.
| Brand | Stryker |
|---|---|
| Model | SERFAS |
| Category | Medical Equipment |
| Language | English |
Alerts to potential risks to patient or physician safety if instructions are ignored.
Highlights procedures to avoid damaging the instrument.
Provides supplementary information for easier maintenance or clarity.
Describes the main unit of the SERFAS Energy System.
Explains how to use the CUT function of the system.
Explains how to use the COAG function of the system.
Describes how to adjust the hand control feature.
Instructions for cleaning the generator unit.
Instructions for cleaning the footswitch.
Instructions and warnings regarding probe cleaning.
Guidance on how to dispose of system components.
Lists the technical specifications for the energy probe.
Lists the technical specifications for the energy generator.
Displays graphs showing output power under various conditions.