SERFAS Energy System
Operation and Maintenance Manual
Endoscopy
Energy System
English, Deutsch, Español, Français, Italiano,
Português, Nederlands, Dansk, Suomi, Svenska,
ǼȜȜȘȞȚțȐ, Polski,
඾ུࢊ , Ё᭛ , ซן૚

Table of Contents

Other manuals for Stryker SERFAS

Questions and Answers

  • K
    Katrina MclaughlinAug 25, 2025
    What to do if Stryker Medical Equipment displays error E4?
    • W
      William NashAug 25, 2025
      If the Stryker Medical Equipment displays error E4, it indicates a low impedance measurement, which signifies a short-circuit. Check the probe for inadvertent contact with metal. If that's not the issue, try replacing the probe or the handpiece cable.
  • S
    Shawn PotterAug 25, 2025
    What to do if Stryker SERFAS Medical Equipment shows error E1?
    • T
      Taylor ElliottAug 26, 2025
      If the Stryker Medical Equipment is showing error E1, it means the device is unable to identify the probe tip. To resolve this, try reconnecting the probe or the handpiece cable. If the problem persists, you may need to replace the probe or the handpiece cable.
  • E
    Edward AllenAug 26, 2025
    What to do if Stryker SERFAS Medical Equipment displays error E2?
    • P
      Patricia HallAug 26, 2025
      If your Stryker Medical Equipment displays error E2, it means the probe or handpiece cable is not being detected when the footswitch pedal is activated. Try reconnecting the probe or the handpiece cable. If that doesn't work, you might need to replace the probe or the handpiece cable.
  • M
    Matthew HudsonAug 27, 2025
    What to do if Stryker SERFAS displays error F1?
    • E
      Evelyn RobinsonAug 27, 2025
      If your Stryker Medical Equipment is displaying error F1, it indicates an RF Current measurement error. To solve this, cycle the power to the unit by turning it off and then on again.
  • S
    Stephanie WagnerAug 27, 2025
    What to do if Stryker Medical Equipment displays error F2?
    • T
      Tom HayesAug 27, 2025
      If the Stryker Medical Equipment displays error F2, it means that there was a hardware shutdown because current, voltage, or power limits were exceeded. To resolve this, cycle the power to the unit (turn it off and then on).
  • D
    Daniel TuckerAug 28, 2025
    What to do if Stryker Medical Equipment displays error F3?
    • O
      Olivia CochranAug 28, 2025
      If your Stryker Medical Equipment displays error F3, it indicates an RF generation or measurement error during the power-on test. Try cycling the power to the unit (turn it off and then on).
  • A
    angelacampbellAug 29, 2025
    What to do if Stryker SERFAS Medical Equipment shows error E7?
    • E
      Ethan SandersAug 29, 2025
      If your Stryker Medical Equipment shows error E7, it means the RF Voltage is too high. You should check the probe for damage and replace it if necessary. Also consider replacing the handpiece cable.
  • M
    Ms. Heather MyersAug 30, 2025
    What to do if Stryker Medical Equipment shows error E8?
    • S
      Sonia LeeAug 30, 2025
      If the Stryker Medical Equipment displays error E8, it indicates the RF Current is too high. You should check the probe for damage and replace it if necessary. Also consider replacing the handpiece cable.
  • K
    Karen SmithAug 23, 2025
    What to do if Stryker SERFAS shows error E6?
    • J
      Joanna ChristensenAug 23, 2025
      If your Stryker Medical Equipment shows error E6, it indicates that the RF Power is too high. You should check the probe for damage and replace it if necessary. Also consider replacing the handpiece cable.
  • T
    Thomas CarrollAug 31, 2025
    What to do if Stryker SERFAS Medical Equipment displays error F5?
    • S
      Steven RyanSep 1, 2025
      If your Stryker Medical Equipment displays error F5, it indicates a COP Timeout error. To solve it, cycle power to the unit (turn the unit off and then on).

Summary

Indications for Use

Contraindications

Warnings & Cautions

Warning

Alerts to potential risks to patient or physician safety if instructions are ignored.

Caution

Highlights procedures to avoid damaging the instrument.

Note

Provides supplementary information for easier maintenance or clarity.

FIRE;EXPLOSION WARNINGS

PRIOR TO SURGERY

DURING SURGERY

AFTER SURGERY

Operating the SERFAS Energy System with the iSwitch

Symbol Definitions

Product Description and Intended Use

The SERFAS Energy Generator

Describes the main unit of the SERFAS Energy System.

The SERFAS Energy Probe

The SERFAS Energy Footswitch

Setting Up the Generator

Setting Up the Probe

Setting Up the Footswitch

Powering On the System

Modes of Operation

CUT MODE

Explains how to use the CUT function of the system.

COAG MODE

Explains how to use the COAG function of the system.

Hand Control Adjustment

Describes how to adjust the hand control feature.

Volume Adjustment

Cleaning

SERFAS Energy Generator

Instructions for cleaning the generator unit.

Footswitch

Instructions for cleaning the footswitch.

Probes

Instructions and warnings regarding probe cleaning.

Equipment Disposal Guidelines

Guidance on how to dispose of system components.

Troubleshooting

Technical Specifications

SERFAS Energy Probe

Lists the technical specifications for the energy probe.

SERFAS Energy Generator

Lists the technical specifications for the energy generator.

RF OUTPUT

Generator Output Graphs

Displays graphs showing output power under various conditions.

Electromagnetic Compatibility

Classifications and Approvals

Warranty

Service and Claims

Stryker SERFAS Specifications