EasyManua.ls Logo

Supmeter BST106-B60S(L) - Weighing, Filtering, and Signal Type Parameters

Supmeter BST106-B60S(L)
39 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
16
No.
Sign
Range
Default
Description
Set
108
Fc
1280
1280
Sampling Frequency [Hz]
1280
109
Ft
0~19
15
Anti-vibration Digital Filter1
Filter1’s 10
0
Digit Set
Set Value
Cutoff Frequency
0
None
1
11.2Hz
2
8.0Hz
3
5.6Hz
4
4.0Hz
5
2.8Hz
6
2.0Hz
7
1.4Hz
8
1.0Hz
9
0.7Hz
Filter1’s 10
1
Digit Set
Set Value
Anti-Vibration Grade
0
Medium
1
High
110
Sb
1~128
1
Weight Display Smooth Filter2
The bigger set value of Filter2 will make the weight
display value more stable with slower response
speed.
111
LC
1~5
1/2/3/4/5
[*]
Weighing Signal Type
1: 19.5EV [0~19.5mV ]
2: 20EA [0~20mA]
3: 2.5V [0~2.5V]
4: 5V [0~5V]
5: 10V [0~10V]
Operating Authorization: Administrator.