6.3.4 Segmenting Span Correction (SEgC)
After doing Zero Calibration and Load Calibration [or Data Calibration], it’s allowed to do
Segmenting Span Correction with loading standard weight on the weigher by 10 Correction Points to get
Span Correction Ratio of 10 linearity segments for higher weighing accuracy.
Main Display Interface
【■】+【►】【 ▲】
【 】+【►】【 ▲】
【 】
M
【 】
Loading ‘Correction Point’s Loading
Weight 1’ on the weigher.
【 】: Display AD Value.
After AD Value is stable, press【 】:
Input ‘Loading Weight 1’:
1~99999 Weight Unit.
[151] Span Correction Ratio 1.
【 】: Save
【 】
Loading ‘Correction Point’s Loading
Weight 2’ on the weigher.
Input ‘Loading Weight 2’:
1~99999 Weight Unit.