EasyManua.ls Logo

Supmeter BST106-B60S(L) - Segmenting Span Correction (Segc)

Supmeter BST106-B60S(L)
39 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
29
6.3.4 Segmenting Span Correction (SEgC)
After doing Zero Calibration and Load Calibration [or Data Calibration], its allowed to do
Segmenting Span Correction with loading standard weight on the weigher by 10 Correction Points to get
Span Correction Ratio of 10 linearity segments for higher weighing accuracy.
Main Display Interface
+】【
+】【
M
Loading ‘Correction Point’s Loading
Weight 1 on the weigher.
: Display AD Value.
After AD Value is stable, press :
Input Loading Weight 1’:
1~99999 Weight Unit.
[151] Span Correction Ratio 1.
: Save
Loading ‘Correction Point’s Loading
Weight 2 on the weigher.
Input Loading Weight 2’:
1~99999 Weight Unit.

Related product manuals