6.3.2 Data Calibration (dAtA)
6.3.2.1【mV】Data Calibration
Input the specification parameter values of loadcell according to the actual configuration of the weighing
system to correct Span Coefficient. If there is no access to get the specification parameter values for Data
Calibration, then it’s necessary to do Load Calibration.
Note:
Total Capacity of Loadcells = Loadcell Capacity × Loadcell Number.
Voltage Ratio = Excitation Voltage on the terminal of controller / Excitation Voltage on the terminal of
loadcells.
The rated excitation voltage for loadcells is DC5V. It’s best to measure the actual voltage value.
Main Display Interface
【■】+【►】【 ▲】
【 】+【►】【 ▲】
【 】
【 】
Input ‘Total Capacity of Loadcells’:
1~99999 Weight Unit.
【 】
【■】: Exit 【 】: Save
M
【 】
Input ‘Output Sensitivity of Loadcell’
[E]: 0.500~5.000mV/V.
Input ‘Voltage Ratio’ [V]:
1.000~2.000.
[105] Span Coefficient value.
【 】
M
Prompt Message:
[19.5EV]:【mV】Data Calibration.