EasyManuals Logo

Symmetricom 5071A Assembly And Service Manual

Symmetricom 5071A
202 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #128 background imageLoading...
Page #128 background image
2. Service
Assembly/Module Diagnostic Trees (Diagnostic Section 3)
112
QUESTION: Is the measured voltage when multiplied by the high
voltage probe calibration factor between 3.3 kV and 4.3
kV?
If Yes: Go to, 3.7.1.2
If No: Remove power, replace A16 and retest.
3.7.1.2 Ion Pump Return Current Test
1 Use the INFO/PUMP menu selection to display the output of the Ion Pump
current monitor circuit.
QUESTION: Was the current displayed between 3 and 5 microamps?
(The high-voltage probe must have an input resistance of 1
Giga.)
If Yes: Go to 3.7.1.3.
If No: Go to 3.7.1.3.1.
3.7.1.3 Ion Pump Return Current Zero Test
1 Power down the instrument, and wait one minute for the A16 module voltages
to bleed down.
2 Disconnect the Ion Pump supply cable from the high-voltage probe and
locate the plug well clear of other conductors.
3 Re-power the instrument, wait 15 seconds, and use the INFO/PUMP menu
selection to display the Output of the Ion Pump current monitor circuit on the
LCD display.
QUESTION: Is the displayed current between 0.0 and 1.0 microamps?
If Yes: Ion Pump high-voltage supply is correct, Go to, 3.7.2.1.
If No: Go to 3.7.1.3.1.

Table of Contents

Other manuals for Symmetricom 5071A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Symmetricom 5071A and is the answer not in the manual?

Symmetricom 5071A Specifications

General IconGeneral
BrandSymmetricom
Model5071A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals