EasyManua.ls Logo

Symmetricom 5071A - Page 129

Symmetricom 5071A
202 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2. Service
Assembly/Module Diagnostic Trees (Diagnostic Section 3)
113
3.7.1.3.1 Ion Pump Current Monitor Diagnosis
1 Power down the instrument, and wait one minute for the A16 module voltages
to bleed down.
2 Disconnect the A16 module from A1 at A1J12.
3 Re-power the instrument, wait 15 seconds, and use the INFO/PUMP menu
selection to display the Output of the Ion Pump current monitor circuit on the
LCD display.
4 Re-connect the A16 module to A1 at A1J12.
QUESTION: Was the displayed current between 0.0 and 1.0
microamps?
If Yes: Remove power, replace A16, and retest.
If No: Go to, 3.7.1.3.2.
3.7.1.3.2 Ion Pump Current Monitor Diagnosis (cont)
1 Power-down the instrument and remove the A2 module from the extender.
2 Re-power the instrument, wait 15 seconds, and use the INFO/PUMP menu
selection to display the Output of the Ion Pump current monitor circuit on the
LCD display.
QUESTION: Is the displayed current between 0.0 and 1.0 microamps?
If Yes: Remove power, replace A2, and re-test.
If No: Remove power, replace A7, and re-test.

Table of Contents

Other manuals for Symmetricom 5071A

Related product manuals