Error message: Low steam
After approx. 40 min. (parameter 68, “PulseTimeOut”),
the pulse was not reached in the pulses before
the sterilisation phase
Steam generator working correctly?
No
Yes
PulsePress 1/2/3 is set higher than
the corresponding steam pressure for
the sterilisation temperature of the
selected program:
Set these parameters lower
- STL (cut-off) triggered
- Steam generator defective
(
see chapter 3.8.1, “Steam generator”)
Steam generator is working,
but does not reach the set value
(
parameter 9/12/15, PlsPress 1/2/3).