7.3.8 Error message: No Gen.Water
Error message: No Gen.Water
No water detected by low electrode
on the generator
- Defective electrode
- Check the steam generator
(
see chapter 3.8.1, “Steam generator”)
7.3.9 Error message: No Tap Water
NoYes
Error message: No Tap Water
Water supply on cooling water connection
greater than 1 bar?
Check the vacuum unit
(
see chapter 3.11, “Vacuum unit”)
Check the water supply