Revision: 2.1 3-67
3.9.2 Maintenance tasks
Testing the safety valve on the steam generator (X series only)
Connect the PC.
Open the Systec service software (see chapter 6, “Systec service
software”).
Call up the window for setting outputs via “I/O”.
Click on Go to manual mode.
Activate the WaterPump output.
The steam generator is now flooded.
Check the steam generator pressure in the control panel display:
The safety valve must open when P
max
is exceeded.
Testing the safety valve on the sterilisation chamber
The functionality of the safety valve must be safeguarded with external
aids.
The safety valve must open when P
max
is exceeded.
Alternative procedure for devices with steam generators:
Unscrew the safety valve on the sterilisation chamber and replace it
with the safety valve on the steam generator.
Follow the same procedure as used to check the safety valve of the
steam generator (see “Testing the safety valve on the steam generator
(X series only)”). Now observe the P
max
of the sterilisation chamber.
After the test, move the safety valves back to their original position.
Alternative procedure (only for devices with support air supply):
Prerequisite: The compressed air supply is able to generate the necessary
pressure.
Attach an external manometer to the sterilisation chamber as a
reference.
Bridge the pressure reducer for the compressed air.
Activate the output in the Service -> Manual output set
menu to allow compressed air to flow.
Check the reference pressure display:
The safety valve must open when P
max
is exceeded.