V/D
S
ERIES
P
ARAMETERS
8-12 Revision: 2.1
No.
Parameter Description Access
level
Range of values Factory
setting
94 WtrChambErr
(Global 19)
For spray cooling or
Eco/Basic devices, for example:
Maximum water intake time for the
sterilisation chamber.
4 50 – 1500 sec. 300 sec.
95 Exh.Time
(Cycle 73)
Parameter e.g. for glass test program:
Defines the exhaust time from the
sterilisation temperature up to EndTemp
“ExhMode” must be set to 1.
4 0 – 1000 min. 0 min.
96 Sleep Power
(Global 20)
Time after which the device will go into
sleep mode if no program is started. The
steam generator is switched off.
0 = Function inactive
4 0 – 999 hours 3 hours
97 PrintAnaBits
(Global 21)
Defines which sensors are printed out by
the integrated printer.
0 = Chamber pressure and temperature
control sensor.
If more sensors should be printed out,
please contact Systec GmbH
4 0 – 9999 0
98 Com Address
(Global 22)
Defines the address of the autoclave in
the RS-485 data bus:
1 = Address undefined
> 1 = Defined address
4 1 – 255 2
99 DoorMode
(Global 25)
Defines whether the door should be
opened after the end of the program (and
which door) on pass-through autoclaves.
0 = Device is not a pass-through
autoclave.
1 = Only one door can be opened at a
time.
2 = Clean room side (back panel): Can
only be opened when the sterilisation
chamber was sterilised again after
the door is opened on the device
side.
3 = Device side (front panel): Can only
be opened when the sterilisation
chamber was sterilised again after
the door is opened on the clean room
side.
4 = If one of the doors was opened, then
the other cannot be opened until the
sterilisation chamber has been
sterilised again.
4 0 – 4 Depends on
the device
Tab. 2: List of parameters for Systec V and D series autoclaves