Revision: 2.1 1/3
M
AINTENANCE CHECKLIST
Please tick off:
Standard device
User asked about faults
1: OK
2: Not OK
3: Repaired/replaced
1
OK
2
Not
OK
3
Repaired
Replaced
Components Inspect/service
Control unit
Control unit Function, display
Door lock
Door Lock, hinge, mechanism
Door seal Wear, cleanliness
Ring switch, door switch
Switching point, function
The ring switch and door switch must be
replaced after 4 years or 2000 cycles.
Pneumatic door opener Function, leaks
Locking ring
- Drive
- Check and lubricate with Teflon
Pressure chamber, hose connections and filter
Pressure chamber
- Corrosion, dirt
- Internal check conforming to German
Ordinance on Industrial Safety and
Health (BetrSichV)
Safety valve on sterilisation
chamber
- Run up to activation pressure
The tolerance is ± 10%
Sterilisation chamber activation
pressure SV:______bar
Safety valve on steam generator
Steam generator activation pressure
SV:______bar
Temperature sensor Damage
Strainer, sieve Clean
Internal compressed air Function, leaks
Pipes and screw connections Seating, leaks
Steam air extractor Function, leaks
Water level electrodes
Function, cleanliness (clean when
necessary)