V/D
S
ERIES
Revision: 2.1 3/3
1
OK
2
Not
OK
3
Repaired
Replaced
Components Inspect/service
Radial fan option
Motor Function, noise
Fan wheel Dirt, free-running
Spray cooling option
Heat exchanger Flow, scaling
Circulation pump Function, noise
Metal hose Leaks, wear
Closed cooling circuit option
Heat exchanger Scaling and free-flowing, clean if
necessary
Circulation pump Function, leaks
Cooling circuit Filling
Lift option
Column Firm seating
Function Travel direction
Arm and basket support Firm seating
Vacuum unit option
Vacuum pump Performance, noise
Sterile air filter Condition, moisture
Vacuum valve Leaks
Pump water valve for vacuum
pump
Function, leaks
Vacuum test Execute
Exhaust filter option
Counter value on exhaust filter
Check
If necessary, replace filter earlier
When replacing filter
Reset counter
Printer option
Printer ribbon Check
Paper supply Check