Page 16 Nano ITC Getting Started Guide
Instrument Specifications
The table found below contains the technical specifications for the Nano ITC instrument.
Table 1: Nano ITC
Technical Specifications
Item/Area Specifications
Dimensions Depth 38 cm (15 in.)
Width 37 cm (14 in.)
Height 32 cm (13 in.)
Weight 20 kg (43 lbs)
Power 100–240 VAC, 3 amps. 50–60 Hz
Operating environmental
conditions
Temperature: 15 to 30°C
Relative Humidity: 5 to 80% (non-condensing)
Installation Category II
Pollution Degree 2
Maximum Altitude: 2500 m (8200 ft)
Emissions class Class A
Temperature range 2 to 80°C
Injection interval 150 s minimum
Response time 15 s
Effective cell volume Standard Volume: 1.0 mL (24K gold or Hastelloy)
Low Volume: 190 µL
Sample volume range Standard Volume: 1200 to 1500 µL
Low Volume: 300 to 700 µL
Injection syringe capacity Standard Volume: 100 or 250 µL
Low Volume: 50 µL
Volume increment 0.114% of the total syringe capacity (0.06, 0.11, or 0.29 µL
respectively for the 50, 100, and 250 µL syringes)
Stirring rate 150 to 400 rpm