Page 30 Nano ITC Getting Started Guide
Figure 14 Syringe with air gap.
Loading the Buret
NOTE: Removal of the buret from the calorimeter is the reverse of this process.
When the syringe is filled with titrant, follow the instructions in this section to load the buret.
The top portion of the buret handle displays a graduated scale with an indicator showing the relative posi-
tion of the syringe plunger during an experiment. The indicator must be in the fully raised position before
installing a loaded syringe into the buret.
1 Insert the plunger and barrel carefully into the rotating shaft of the buret assembly. (See the figure below.)
Figure 15 Plunger and barrel.
CAUTION: To avoid possible damage, do not use the 100 or 250 µL syringes with the Nano ITC
Low Volume instrument.

Table of Contents

Questions and Answers:

TA Instruments Nano ITC 5300 Specifications

General IconGeneral
BrandTA Instruments
ModelNano ITC 5300
CategoryMeasuring Instruments
LanguageEnglish

Summary

Notes, Cautions, and Warnings

Low Voltage Directive and Safety

Covers safety compliance and warnings for electrical safety.

Lifting the Instrument

Warnings about moving the instrument.

System Components

Instrument Specifications

Installing the Nano ITC

Instrument Installation Steps

Procedures for setup and recommendations.

Choosing a Location for the Nano ITC

Guidelines for ideal placement.

Power Requirements and Connection

Details power source, surge protection, and connection.

Setting Up the Nano ITC

Basic setup steps.

Connecting Cables and Cords

How to connect power and USB cables.

Starting the Nano ITC

Steps to power on and launch software.

Use, Maintenance, & Diagnostics

Calibrating Chemicals

Discusses calibration methods and examples.

Experiment Parameters for Calibration

Lists parameters for calibration.

Calibration Factor Input Location

Shows where to input calibration factor in software.

Preparing and Degassing Solutions

Detailed steps for preparing samples and buffers.

Starting the Experiment

Steps to begin a titration experiment.

Cleaning the Nano ITC

Procedures for cleaning the instrument and its components.

Troubleshooting the Nano ITC

Common issues and solutions.

Buret Position Functions

Related product manuals