Nano ITC Getting Started Guide Page 23
Chapter: 3
Use, Maintenance, & Diagnostics
Overview
A typical Nano ITC experiment involves the following:
Preparing and degassing the solutions
Preparing the sample and reference cells
Mounting the buret
Running the baseline
Performing an analysis
Cleaning the calorimeter
Each step is briefly described here. Additional information is provided in the online help supplied with the
software program. It is assumed that you are familiar with standard laboratory procedures and techniques.
It is critical that the Nano ITC cells be cleaned immediately at the end of each experiment. The calorimeter
can be left idle for up to two days with water-filled cells at 25°C, when not performing experiments. When
the instrument is expected to remain idle longer than five days, clean and empty the cells.
The Nano ITC can be calibrated using standardized chemical reactions. If suitable reagents are not
available, then as an alternate option perform the calibration using electrical pulses.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TA Instruments Nano ITC 5300 and is the answer not in the manual?

TA Instruments Nano ITC 5300 Specifications

General IconGeneral
BrandTA Instruments
ModelNano ITC 5300
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals